METHOD PROCEDURES
67
67
SULFATE
• Add one packet of HI93751-0 Sulfate Reagent.
Replace the plastic stopper and the cap. Invert gently for
1 minute (about 30 inversions).
• Insert the cuvette into the holder and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or wait 5 minutes
and press Read. When the timer ends the meter will perform the reading. The instrument displays
the concentration in mg/L of sulfate (SO
4
2-
).
INTERFERENCES
Interference may be caused by:
• Chloride above 40000 mg/L
• Calcium above 20000 mg/L CaCO
3
• Magnesium above 10000 mg/L MgCO
3
• Silica above 500 mg/L SiO
2
• Color or suspended matter, filter the sample prior to analysis
• Organic matter in large amounts may impede the precipitation of barium sulfate