EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>HI97711

Hanna Instruments HI97711 User Manual

Hanna Instruments HI97711
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
12
GENERAL OPERATIONS
4. Insert the A ZERO CAL Check Cuvette A then press Next to continue. The Please
wait... message will be displayed during the measurement.
5. Insert the HI97701B CAL Check Cuvette B then press Next to continue. The
Please wait... message will be displayed during the measurement.
6. When the CAL Check is complete, the display will show one of the following
messages and the value obtained during the measurement:
“PASSED“: The measured value is within the accuracy specification, no user
calibration is required.
OUT OF SPECIFICATION and Calibrate is available: The measured value is
near the expected value. To update the user calibration, press Calibrate. Press
Accept to confirm or Cancel to return to the previous screen.
OUT OF SPECIFICATION: A user calibration is not allowed, the measured
value is outside of the tolerance window. Check the certified value, expiration
date and clean the outside of the cuvette. Repeat the CAL Check procedure. If this
error continues, contact your nearest Hanna Instruments Customer Service Center.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI97711 and is the answer not in the manual?

Hanna Instruments HI97711 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI97711
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals