METHOD PROCEDURE
8.3. TOTAL CHLORINE (POWDER REAGENT)
Note: Free and Total Chlorine have to be measured separately with fresh unreacted
samples, following the related procedures, if both values are desired.
REQUIRED REAGENTS
Code Description Quantity
HI93711-0 Total Chlorine Reagent 1 packet
REAGENT SETS
HI93711-01 Total Chlorine Reagent - 100 tests
HI93711-03 Total Chlorine Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Select the Total Chlorine (Powder) method using the procedure described in the
METHOD SELECTION section.
Note: If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press Measure and follow the messages on the screen.
Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI97711 and is the answer not in the manual?

Hanna Instruments HI97711 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI97711
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals