METHOD PROCEDURE
• Remove the cuvette.
• To an empty cuvette add 3 drops of
HI93701A-T Total Chlorine Reagent A,
3 drops of HI93701B-T Total Chlorine
Reagent B, and 1 drop of HI93701C-T
Total Chlorine Reagent C.
• Swirl gently to mix.
• Add 10 mL of unreacted sample (up to the
mark). Replace the plastic stopper and the
cap. Shake gently to mix.
• Insert the cuvette into the holder and ensure
that the notch on the cap is positioned securely
in the groove.
• Press Read. The display will show a 2 minute 30 second countdown prior to the
measurement. To skip the timer, press Read twice. When the timer ends, the
meter will perform the reading. The instrument displays the results in mg/L of
chlorine (Cl
2
).
×3 ×3 ×1
10 mL