Hanna Instruments HI991300 - Specifications for HI991301; Range, Resolution, Accuracy; Calibration and Compensation; Probe, Battery, Environment

Hanna Instruments HI991300
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
SPECIFICATIONS
HI991301
Range*
‑2.00 to 16.00 pH / ‑2.0 to 16.0 pH
±825 mV (pH‑mV)
0.00 to 20.00 mS/cm**
0.00 to 10.00 ppt
‑5.0 to 105.0 °C / 23.0 to 221.0 °F
Resolution
0.01 pH / 0.1 pH
1 mV
0.01 mS/cm
0.01 ppt (g/L)
0.1 °C / 0.1 °F
Accuracy
@ 25°C/77°F
±0.02 pH / ±0.1 pH
±1 mV (pH‑mV)
±2% F.S. (EC/TDS)
±0.5 °C / ±1.0 °F Temperature
Temperature
compensation
pH ‑ Automatic
EC/TDS ‑ Automatic, with β selectable from
0.0 to 2.4 %/ °C (0.1 increments)
pH calibration
Automatic, 1 or 2 points choice between
2 sets of buffers (standard: 4.01, 7.01,
10.01 or NIST: 4.01, 6.86, 9.18)
EC/TDS
calibration
Automatic, one‑point at 12.88 mS/cm or
6.44 ppt (CONV=0.5) or
9.02 ppt (CONV=0.7)
TDS conversion
factor
Selectable from 0.45 to 1.00
(0.01 increments)
Probe (included)
HI12883 pH/EC/TDS/temperature sensor,
DIN connector and 1 m (3.3’) cable
Battery type/life
1.5V AAA (3 pcs.)
approx. 600 hours of continuous use
Auto‑off User selectable: after 8 min, 60 min or disabled
Environment 0 to 50 °C (32 to 122 °F) RH max. 100%
Meter dimensions 154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight
(with batteries)
196 g (6.91 oz.)
Casing ingress
protection rating
IP67
* the HI12883 is limited to be used from 0 to 13 pH and from 0 to
50 °C (32 to 122 °F)
** LCD displays mS for mS/cm