EasyManua.ls Logo

Hanna Instruments HI991300 - Meter Setup Options; Setup Mode Navigation; Temperature Unit Selection; Auto-off and Beep Tone Settings

Hanna Instruments HI991300
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
METER SETUP
Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep confirmation tone, the type of pH buffer set,
the pH Resolution, enable/disable calibration information
display, temperature compensation factor for EC and TDS
conversion factor. To enter Setup mode press and hold
button until “EC STD“ and tag is replaced by
SETUP“ and tag. Release the button.
• “TEMP“ is displayed on the third
LCD line with the current temperature
unit (e.g. “TEMP °C“), for °C/°F
selection, use the SET button. After
the temperature unit has been
selected, press
to confirm and to
enter the “A-OFF“ selection.
Use the SET button, to navigate through
the auto‑off choices: 8 minutes (“8”,
default value), 60 minutes (“60”) or
disabled (“---”). Press
to confirm
and to enter the “BEEP“ selection.
To switch the beep tone ON or OFF,
press the SET button; press
to
confirm and to enter the calibration
buffer selection “pH 7.01 BUFF“.
To change the buffer set, the meter
will show the current buffer set: “pH
7.01 BUFF” (for standard buffer
set: 4.01/7.01/10.01) or “pH
6.86 BUFF” (for NIST buffer set:
4.01/6.86/9.18). Change the set with the SET button.
Press
to confirm and to enter pH resolution selection
RESOL“.
• To change the pH measurement
resolution between “0.1“ and “0.01
use the SET button; then press to
confirm and to enter electrode calibration
information “INFO“ selection.
METER SETUP

Related product manuals