Chapter 5 Specification
The general and measurement precision specification of LCR are described below.
5.1 General Specification
Hantek1832C
Hantek1833C
Main parameter: L/C/R/Z Secondary parameter:
X/D/Q/θ/ESR
Series, parallel
Manual, automatic
Fast(4 times/s), medium(2 times/s), slow(1 times/s)
3-terminal, 5-terminal
Short circuit, Open circuit
Type C(virtual serial port)
100Hz, 120Hz, 400Hz,
1kHz, 4KHz, 10kHz,
40kHz
100Hz, 120Hz, 400Hz,
1kHz, 4KHz, 10kHz,
40kHz, 50KHz, 75KHz,
100kHz
0.6Vrms
0.3Vrms, 0.6Vrms
100Ω
Resistance: 0.25%
Capacitance: 0.4%
L: 0-2000H C: 0-20mF R: 0-20MΩ
5.2 Accuracy
Notes:
Environment temperature: 20
± 2
, humidity: 75% R.H;
Preheat the instrument for at least 30 minutes before the test;
Test at the test notch on the end face of the instrument;
It is better to conduct open and short circuit correction before the test;
Measure with the recommended equivalent mode;
The percentage in terms of error indicates:
± (% of the reading + last digit)
If the actual measurement of the instrument and the display exceeds the
scope specified in the table, the accuracy of the excessive part will not be
given;
The meaning of the subscript
S- series equivalent; p- parallel equivalent; e- accuracy
Some parameters cannot be given in the data table, and it can only be