Basic Operation
DSO4000 Series Digital Storage Oscilloscope 39
2.13 UTILITY System
Push the UTILITY button to display the Utility Menu as follows.
Options
Comments
System Info
Display the software and hardware versions, serial number and some other
information about the oscilloscope.
Update Program
Insert a USB disk with upgrade program and the disk icon at the top left
corner is highlighted. Press the Update Program button and the Software
Upgrade dialog pops up. Push F6 to upgrade or push F2 to cancel.
Save Waveform
Insert a USB disk and the disk icon at the top left corner is highlighted. Click
on this icon and you can see the waveform pause for a while, being saved.
You can find the saved waveform data under the Hantek_x folder in the USB
disk. Here X represents how many times you press the button. Each press
generates a corresponding folder. For example, press once and a folder
Hantek_1 is generated; press twice and two folders Hantek_1, Hantek_2
are generated.
Self Calibration
Press this option and the Self Calibration dialog pops up. Press F6 to
perform the self calibration. Press F4 to cancel.
Keepad Beep
Turn on or turn off key beep.
GUI Language
Set the language
GUI color
Set the GUI color, bule, black, green, pink, yellow.
Time
Set the time setting.
Sys Status
Display the system status.
Pass/Fail
Pass/Fail function
Recorder
Turn on record function.
Filter
Filter setting.
Print Cfg
Set print configuration
Option
Bode Assiant:
Wave: Save waveform as .hws file to Flash, SD or USB device.
DDS: Enable waveform generator function.
DVM:
Tip: Press any menu button on the front panel to remove the status display and enter a
corresponding menu.
2.13.1 Firmware Update
This series of oscilloscope can upgrade the software by the USB flash disc, which needs about 5
minutes.
The firmware is upgraded by the following steps:
1. Plug the USB flash disc in which a firmware program is saved in USB Host interface
on the front panel of the oscilloscope.
2. Press [UTILITY] to skip to the “UTILITY” menu.
3. Press “Firmware Update”.