Instructions for the metrological check
Every Veroval
®
monitor has been carefully tested by HARTMANN for
measurement accuracy and has been developed with a view to a long service
life. We recommend carrying out a metrological check every 2years for
devices in professional use, for example, in pharmacies, medical practices or
hospitals. Please also observe the national regulations specified by the legislator.
Metrological checks should only be carried out by the competent authorities or
authorised maintenance providers at the user‘s expense.
Instructions for calibration mode
A functional check of the device can be carried out on a person or using
a suitable simulator. A metrological check involves testing for leaks in the
pressure system and potential deviations in the pressure reading. Remove at
least one battery to switch to calibration mode. Now keep the START/STOP
button pressed and re-insert the battery. Keep it pressed for a few more
seconds until the display shows a flashing ‘ ‘. Then release the button and two
zeros ‘ ‘ appear, shown on top of one another, in the display. Instructions on
the metrological check will be supplied by HARTMANN upon request to the
competent authorities or authorised maintenance providers.
Disposal information
To protect our environment, empty batteries must not be disposed of in
household waste. Please comply with the relevant waste disposal regulations
or use public collection points.
This product is subject to the European Directive 2012/19/EU on Waste
Electrical and Electronic Equipment and is marked accordingly. Never
dispose of electronic equipment in your household waste. Please
obtain information about your local regulations on the proper disposal
of electrical and electronic products. Proper disposal protects the
environment and human health.
IFU_Veroval_dc_SK6_0307662_2020-02-10.indd 16 27.02.20 13:08