14. Power supply
Model no.: LXCP12-006060BEH
Input: 100 – 240V~, 50 60Hz, 0.5A max.
Output: 6V DC, 600mA, only in conjunction with the
Veroval
®
duo control blood pressure monitor
Manufacturer: Globalcare Medical Technology Co., Ltd.
Protection: The device is double-insulated and has a primary
fuse that disconnects the device from the mains in
the event of a fault.
Polarity of the DC voltage connection
Protective insulation/protection class2
Housing and protective
covers:
The power supply unit housing prevents touching
contact being made with parts that are carrying or
can carry a current (fingers, needles, test hooks).
The user must not touch the patient and the outlet
plug of the AC/DC power supply unit at the same
time.
Legal requirements and guidelines
The device also conforms to the requirements of the European standard
EN1060: Non-invasive sphygmomanometers – Part 1: General requirements
and Part 3 – Supplementary requirements for electro-mechanical blood
pressure measuring systems; as well as the standard EN80601-2-30.
Clinical testing of the measuring accuracy was carried out in accordance
with EN1060-4 and EN81060-2. The requirements of ANSI/AAMI test
protocol SP10-1992 are also fulfilled.
Subject to change without prior notice
IFU_Veroval_dc_SK6_0307662_2020-02-10.indd 37 27.02.20 13:09