13. Technical data
Product description: Automatic upper arm blood pressure monitor
Model: Veroval
®
duo control
Type: DC3-18
Measuring method: Korotkoff and Oscillometric
Display range: 0300mmHg
Measuring range: Systole (SYS): 50–250mmHg
Diastole (DIA): 40–180mmHg
Pulse: 30–199beats/minute
Display unit: 1mmHg
Technical measurement
accuracy:
Cuff pressure: ± 3mmHg
Pulse: ± 5% of the displayed pulse rate
Clinical measuring
accuracy:
Corresponds to the requirements of DINEN1060-4
and DINENISO81060-2; Korotkoff validation
method: PhaseI (SYS), PhaseV (DIA)
Operating mode: Continuous operation
Nominal voltage: DC 6V
Power supply: 4× 1.5V alkaline manganese Mignon (AA/LR06)
batteries or the optional Veroval
®
power supply
unit
Battery capacity: approx. 950measurements
Protection against electric
shock:
Internally powered medical electrical (ME)
equipment (only when batteries are used). Applied
part: Type BF
= ClassII ME device (when using the Veroval
®
power supply unit)
Protection against
harmful ingress of water
or solid materials:
IP20 (not protected against moisture,
protected against solid objects ≥ 12.5mm)
Inflation pressure: min. 140mmHg
Comfort Air Technology: Individual inflation pressure depends on systolic
blood pressure +30mmHg