13. Technical data
Product description: Automatic upper arm blood pressure monitor
Model: Veroval
®
duo control
Type: DC3-18
Measuring method: Korotkoff and Oscillometric
Display range: 0300 mmHg
Measuring range: Systole (SYS): 50–250 mmHg
Diastole (DIA): 40–180 mmHg
Pulse: 30–199 beats/minute
Display unit: 1 mmHg
Technical measurement
accuracy:
Cuff pressure: ± 3 mmHg
Pulse: ± 5% of the displayed pulse rate
Clinical measuring
accuracy:
corresponds to the requirements of DIN EN 1060-4
and DIN EN ISO 81060-2; Korotkoff validation
method: Phase I (SYS), Phase V (DIA)
Operating mode: Continuous operation
Nominal voltage: DC 6V
Power supply: 4× 1.5V alkaline manganese Mignon (AA/LR06)
batteries or the optional Veroval
®
power supply
unit
Battery capacity: approx. 950 measurements
Protection against electric
shock:
Internally powered medical electrical (ME)
equipment (only when batteries are used). Applied
part: Type BF
= Class II ME device (when using the Veroval
®
power supply unit)
Protection against
harmful ingress of water
or solid materials:
IP20 (not protected against moisture, protected
against solid objects ≥ 12.5 mm)
Inflation pressure: min. 140 mmHg
Comfort Air Technology: individual inflation pressure depends on systolic
blood pressure +30 mmHg
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 33 17.10.18 18:04

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hartmann Veroval DUO CONTROL and is the answer not in the manual?

Hartmann Veroval DUO CONTROL Specifications

General IconGeneral
BrandHartmann
ModelVeroval DUO CONTROL
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals