13. Technical data
Product description: Automatic upper arm blood pressure monitor
Model: Veroval
®
duo control
Type: DC3-18
Measuring method: Korotkoff and Oscillometric
Display range: 0300 mmHg
Measuring range: Systole (SYS): 50–250 mmHg
Diastole (DIA): 40–180 mmHg
Pulse: 30–199 beats/minute
Display unit: 1 mmHg
Technical measurement
accuracy:
Cuff pressure: ± 3 mmHg
Pulse: ± 5% of the displayed pulse rate
Clinical measuring
accuracy:
corresponds to the requirements of DIN EN 1060-4
and DIN EN ISO 81060-2; Korotkoff validation
method: Phase I (SYS), Phase V (DIA)
Operating mode: Continuous operation
Nominal voltage: DC 6V
Power supply: 4× 1.5V alkaline manganese Mignon (AA/LR06)
batteries or the optional Veroval
®
power supply
unit
Battery capacity: approx. 950 measurements
Protection against electric
shock:
Internally powered medical electrical (ME)
equipment (only when batteries are used). Applied
part: Type BF
= Class II ME device (when using the Veroval
®
power supply unit)
Protection against
harmful ingress of water
or solid materials:
IP20 (not protected against moisture, protected
against solid objects ≥ 12.5 mm)
Inflation pressure: min. 140 mmHg
Comfort Air Technology: individual inflation pressure depends on systolic
blood pressure +30 mmHg