37
Electromagnetic in-
terference immunity
tests
IEC 60601 test
level
Compliance level Electromagnetic environment
– guidance
Magnetic field at
supply frequency
(50 Hz) as per
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of those found
in a typical commercial or
hospital environment.
Table 3
For NON-LIFE-SUPPORT MEDICAL-ELECTRICAL (ME) DEVICES and ME
SYSTEMS
Manufacturer declaration and guidance – Electromagnetic immunity
The Veroval
®
duo control blood pressure monitor is intended for use in one of the
electromagnetic environments specified below. The customer or user of the Veroval
®
duo
control blood pressure monitor should assure that it is used in such an environment.
Electromagnetic
interference immunity
tests
IEC 60601 test level Compliance
level
Electromagnetic environment
– guidance
Portable and mobile RF
communications equipment
should be used no closer
to any part of the Veroval
®
duo control blood pressure
monitor, including its cables,
than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter. Recommended
separation distance:
conducted RF
disturbances as per
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
radiated RF interference
in accordance with
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz