3. Malfunction or damage caused by fire, salt, poisonous gas, earthquake, hurricane, flood, abnormal
electric voltage or any other natural disaster. For all the malfunctions and damage due to above
reasons, the manufacturer can offer repair but charge for the cost.
16. EMC Information
Note:
The infusion pump conforms to EMC relevant requirements of IEC 60601-1-2 standard;
Users should install and use the product according to the EMC information provided by random
documents;
Since portable and mobile RF communication device may affect the performance of infusion pump,
strong electromagnet interference should be avoided, for example, do not use it close to mobile
phone or microwave;
See attachments for the guidance and manufacturer’s declaration.
Warning:
The infusion pump should not be put close to or stacked on other devices. If it has to be put close
to or stacked on other device, it should be observed and verified that if it can operate normally
under its application configuration;
Class A device is planned to be used in industrial environment. Due to the conducted disturbance
and radiated disturbance of infusion pump, it may be difficult to ensure EMC in other environment;
To use accessories and cables not provided by infusion pump manufacturer, which may increase
emission or reduce interference immunity of the pump.