8.4 System Setting
Select “Settings” in the main menu, press OK key to enter.
Press key to select requested parameter, press OK key to make it editable, input
required value, then press OK key to save the setting. All the setting parameters can be saved
automatically after turn off the device, and storage time is 5 years.
8.4.1 Set alarm sound volume
Select Settings” in the main menu, press OK key to enter. Press OK key for ‘Sound’, press or
key to select required sound volume (low, medium, high). Then press OK key to save the setting
and exit.
Alarm sound volume is not lower than 45dB when setting as low level, and not higher than 65dB when
setting as high level.
8.4.2 KVO rate setting
KVO means to run the infusion at the preset low speed in order to keep vein open. Press OK to save
after setting
8.4.3 Occlusion level setting