What to do if the main power switch is on, and the doors are closed, but the built-in platelet agitator is not agitating (PC4200i) in my Helmer Accessories?
- EElizabeth HaynesSep 23, 2025
If the main power switch is on, and the doors are closed, but the built-in platelet agitator is not agitating, confirm the hinge cams are not damaged and replace if necessary. Also, confirm the door gaskets are sealing completely and replace the door gasket if necessary. Confirm that all circuit breakers are seated and push the circuit breaker to reset it if necessary.