EasyManua.ls Logo

HemoCue WBC DIFF - Page 39

HemoCue WBC DIFF
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
8. When the blood drop is large enough, fill the microcuvette in one continuous process.
Do NOT refill!
NOTE: Make sure that the microcuvette is filled from the tip, forming an angle (about
45 °) with the blood drop.
9. Wipe off excess blood from the outside of the microcuvette with a clean, lint-free wipe.
Do not touch the open end of the microcuvette.
10. Look for air bubbles in the filled microcuvette. If present, discard the microcuvette and
fill a new microcuvette from a new drop of blood.
NOTE: Make sure that the microcuvette is lled correctly as an improper lling angle might
cause air bubbles.
NOTE: If a second sample is to be taken, it is important that this is done after the measure-
ment of the rst sample is completed. Wipe away the remains of the drop of blood and ll
the second microcuvette from a new drop of blood as per steps 8–10 above.
NOTE: HemoCue WBC DIFF is not validated for heel stick.

Related product manuals