HemoCue WBC DIFF - Method Comparison Study - Capillary Samples Direct from Finger

HemoCue WBC DIFF
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
75
Method Comparison Study - Capillary Samples Direct from Finger
Results from a method comparison study performed on the HemoCue WBC DIFF system and the Sysmex XS-1000i are shown for total
WBC count in Figure 1, for neutrophil count in Figure 2 and for lymphocytes in Figure 3. Capillary sample for Sysmex was taken in a
microtube.
Coefficients:
y = 0.33 + 0.88x
n = 96
r
2
= 0.867
r = 0.931
y = x
Sysmex XS-1000i, Neutrophils (x10
9
/L)
HemoCue WBC DIFF, Neutrophils (x10
9
/L)
Coefficients:
y = 0.47 + 0.94x
n = 103
r
2
= 0.859
r = 0.927
y = x
HemoCue WBC DIFF total WBC (x10
9
/L)
Sysmex XS-1000i, total WBC (x10
9
/L)
HemoCue WBC DIFF, Lymphocytes (x10
9
/L)
Sysmex XS-1000i, Lymphocytes (x10
9
/L)
Coefficients:
y = 0.27 + 1.10x
n = 96
r
2
= 0.807
r = 0.898
y = x
Figure 1 Total WBC Figure 2 Neutrophils Figure 3 Lymphocytes