Patient Test stands for performing tests on samples from patients.
1. Start the analyzer as described in Start-Up section. Place the cuvette moving arm in
loading position.
2. The display will show the main menu. Take a microcuvette from the package
(HemoCue WBC DIFF Microcuvette or HemoCue WBC Microcuvette). For single
packaged microcuvettes; open the inner wrapping and take out the microcuvette.
3. Press the button below Patient Test symbol.
4. Enter required data (selected in settings) e.g. Operator ID, Patient ID, Lab ID, and Site
ID. To enter data see Data Entry section.
5. If data has been entered the data confirmation window is shown. Press right button to
confirm. To change information, press left button until the desired window is shown.