7. If data has been entered the data confirmation window is shown. Press right button to
confirm. To change information, press left button until the desired window is displayed.
8. Insert cuvette symbol is shown.
9. Take a sample according to Blood Collection Venous Sample and Control Material sec-
tions. Place the microcuvette into the cuvette holder (9a) and start measurement as soon
as possible but no later than 1 minute after filling the microcuvette by gently pushing
the cuvette holder to its measuring position (9b). It will automatically slide into the cor-
rect position and the measurement starts.
10. During measurement the measuring window is displayed.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HemoCue WBC DIFF and is the answer not in the manual?

HemoCue WBC DIFF Specifications

General IconGeneral
Sample Volume10 µL
Reportable Range0.3–30.0 x 10^9/L
Measurement ParametersNeutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils
Sample TypeCapillary or venous whole blood
Operating Temperature15-30°C
ConnectivityUSB, Ethernet
CalibrationFactory calibrated
Measurement TimeApproximately 3 minutes
DisplayColor LCD