59
Clean the cuvette holder after each day of use.
1. Turn off the analyzer. Place the cuvette moving arm in loading position.
2. Remove cuvette holder by lifting it straight up.
3. Clean the cuvette holder with alcohol (20-70%) or mild detergent.
NOTE: Do not autoclave.
If the optical parts are stained an error code is displayed. Clean according to
Maintenance/Optical Parts section.
4. Wait 15 minutes before replacing the cuvette holder.
5. Place the cuvette moving arm in loading position before replacing the cuvette holder.
Clean the cover with alcohol (20-70%) or a mild detergent.
Only the WBC DIFF cuvette holder is to be used with the WBC DIFF analyzer.
The WBC DIFF cuvette holder is marked WBC DIFF.