Specimen Collection and Preparation
Capillary or venous whole blood may be used. EDTA anti-coagu-
lant should be used. Do not dilute sample.
Venous blood samples and capillary blood samples in EDTA
tubes shall be stored at room temperature 18–30°C, (6486°F).
Stability: venous samples 8h, capillary samples 4h.
Materials Required
HemoCue WBC DIFF Analyzer
HemoCue WBC DIFF Microcuvettes and/or HemoCue WBC
Microcuvettes
Lancet (capillary samples)
Pipette or other transfer device (venous samples)
Lint-free tissue (non-fraying)
Quality Control
When the HemoCue WBC DIFF Analyzer is started, a quality
control (self-test) is automatically performed to verify perfor-
mance. If the test fails, an error code will be displayed.
For each measurement, quality controls are made on:
The HemoCue WBC DIFF Analyzer.
The HemoCue WBC DIFF Microcuvettes or WBC Microcu-
vettes.
The sample.
The handling of microcuvettes and sample.
No additional quality controls performed by the operator are
required for functionality verification.