Clean and Disinfect
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6) 47
Clean and Disinfect
We recommend that you clean and disinfect the Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
between patient use and regularly during extended patient stays. Refer to your facility’s cleaning and disinfection
policies, and follow the guidelines below.
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could cause injury or equipment damage.
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment. Failure to follow facility protocol could
cause death or serious injury.
SHOCK HAZARD:
Failure to unplug the bed from its power source before you clean or disinfect the bed could cause injury or
equipment damage.
SHOCK HAZARD:
Do not expose the bed to excessive moisture. Injury or equipment damage could occur.
CAUTION:
Do not steam clean or power wash the bed or mattress. Pressure and excessive moisture can damage the
mattress and the protective surfaces of the bed and its electrical components.
CAUTION:
Do not use harsh or abrasive cleansers, solvents, or scouring pads. Equipment damage could occur.
CAUTION:
Make sure the bed frame and mattress are dry before you put the mattress on the bed. Failure to do so
could cause equipment damage.
NOTE:
Clean and disinfect surfaces with nano Ag+® Technology as you would other surfaces. This product is not
intended to protect users or others against bacteria, viruses, germs, or other disease organisms. Always disinfect
Hill-Rom surfaces thoroughly after each use according to facility protocol. Refer to “Clean” and “Disinfect” below.