4.1 Lamp
Chapter 4: Removal, Replacement, and Adjustment Procedures
Prima™ Surgical Light Service Manual (man237ra) Page 4 - 3
4
4.1 Lamp
Tools required: None
Removal
SHOCK HAZARD:
To minimize the risk of electrical shock or damage to equipment,
disconnect all electrical power to the system before working on it.
Failure to do so could result in personal injury or equipment damage.
1. Perform the following:
a. Locate the involved building standard/emergency circuit breaker panel.
b. Set the involved circuit breaker (A) to OFF (see figure 4-1 on page
4-3).
Figure 4-1. Standard or Emergency Circuit Breaker Box
c. Lockout and tag out the breaker.
m237
011