56 The Vest
®
Airway Clearance System, Model 205
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
Use the disinfectant as specified in the disinfectant manufacturer's
instructions. Do not attempt to disinfect or sterilize the disposable
garment for reuse with more than one patient.
Maintenance
Only authorized persons should service The Vest
®
Airway
Clearance System, Model 205. Service by unauthorized persons
could cause personal injury or equipment damage.
Minimal routine maintenance and periodic cleaning are necessary for
The Vest
®
Airway Clearance System, Model 205.
Facilities should do these tests and examinations annually:
• Unplug the Air Pulse Generator from its power source.
• Examine the overall condition of the unit for damage or missing
parts.
• Examine the Power Cord and connector for cuts, scrapes, or other
damage.
• Do electrical safety tests in accordance with facility protocols.
• Clean and disinfect the unit (see “Cleaning” on page 54).
• Connect the Air Pulse Generator to a disposable garment and to an
appropriate power source. Make sure the unit is operational, and that
all functions operate correctly.