Welch Allyn
®
Spot Vital Signs 4400
Quick reference
Power button
• Located on the device housing
• Powers up the device
• When the device is on, opens pop-up dialog with controls
to power down, enter sleep mode, or cancel out of the
power menu
Power down
1. Touch the power button.
2. Touch Power down.
Battery status
Charging Battery removed or
not holding a charge
Approximate
operating time
remaining
NIBP measurement
Stat/Stop Stat average/Cancel/Skip
START STOP
CANCEL
START
AVERAGE
SKIP
Stat an NIBP averaging program
The averaging program takes consecutive NIBP measurements within a
congurable amount of time. When the program is complete, the average
measurement is displayed in the Home tab.
1. Place the proper cu around the patient’s bare upper arm.
2. On the Home tab, touch Start Average.
3. If necessary, touch Skip to stop the current measurement. The
measurement restarts at the end of the countdown.
4. When averaging completes, touch Save to save the patient data, or touch
Clear to discard the data.
Results of an NIBP averaging program
NIBP AVERAGED
SYS/DIA mmHg ( 76 MAP )
111/62
1. 120/80
2. 123/82
3. 110/64
4. 111/62
5. 108/61
6. 105/59
TOTAL READINGS AVERAGED: 4
Displays the average of the readings.
Displays “NIBP AVERAGED.”
Displays the readings taken by the program. A reading with a line
through it is excluded from the average.
Note
To retain an NIBP average, touch Save.