EN-8
Inspection Before Measurement
Inspection Before Measurement
Verify that the instrument operates normally to
ensure that no damage occurred during storage
or shipping. If you nd any damage, contact your
authorized Hioki distributor or reseller.
If damage is suspected, check the section below
before contacting your authorized Hioki distributor
or reseller.
1
Check that the test lead is not broken.
Replace with the speci ed L9208 Test Lead.
2
Check that the resistance measurement and
continuity test operates normally.
Have the instrument repaired by the your
authorized Hioki distributor or reseller. The
instrument may have been subject to a voltage
of greater than 600 V during resistance
measurement or continuity testing.
3
Check that the battery voltage is not low.
Replace the batteries.

Related product manuals