    Zoe EllisAug 16, 2025
    Why are setting commands rejected on my Hioki 3169-20 Measuring Instruments?
    • D
      dleeAug 16, 2025
      The settings cannot be changed while time-series measurement is on standby or in progress, or the measurement data is being held.
  • J
    Jorge StoneAug 18, 2025
    Why does the response message to a query differ from the display of the Hioki 3169-20/21 Measuring Instruments?
    • M
      Mary TurnerAug 19, 2025
      A response message is created after the Hioki 3169-20/21 has received a query. The message, therefore, may differ from the display when the PC reads the response message.
  • B
    bmcculloughAug 22, 2025
    What to do if the number of read data is too small with Hioki 3169-20 Measuring Instruments?
    • H
      howard71Aug 22, 2025
      If the data contains a comma, try using a LINE INPUT statement.
  • J
    John WeberAug 25, 2025
    How to fix communication failure with RS-232C on Hioki 3169-20?
    • G
      George PetersAug 26, 2025
      To resolve the communication failure, ensure that: * The message terminator (delimiter) of the PC is set correctly. * The PC settings match the RS-232C connection conditions set on the 3169-20/21.
  • A
    Alicia GarciaAug 27, 2025
    What to do if the RS-232C bus hangs when reading data using BASIC INPUT statement on Hioki 3169-20 Measuring Instruments?
    • E
      Elizabeth SaundersAug 28, 2025
      When attempting to read data using a BASIC INPUT statement and the RS-232C bus hangs, ensure that: * A query is transmitted before every INPUT#(ENTER). * None of the transmitted queries have resulted in an error.
  • C
    Charles GonzalezAug 31, 2025
    How to troubleshoot Hioki 3169-20 Measuring Instruments when the RS-232C has stopped working completely?
    • C
      Connie RamirezAug 31, 2025
      To troubleshoot a completely non-functional RS-232C connection on your Hioki Measuring Instruments, check the following: * Are the cables properly connected? * Are all the devices powered on? * Are the correct cables being used? * Is the device to be connected to the RS-232C set to a PRINTER? If so, change the setting to a PC.
  • J
    Joseph JohnsonSep 3, 2025
    What to do if sending several queries produces only one response on Hioki Measuring Instruments?
    • A
      Andrew TorresSep 3, 2025
      When several queries produces only one response, consider the following: * Has an error occurred? * Send the queries one at a time, and read the responses individually. When you want to read them in all at once, try doing so by putting them all on one line separated by the message separator character.
  • T
    troweSep 5, 2025
    What to do if nothing happens after transmitting a command to Hioki 3169-20 Measuring Instruments?
    • D
      Danielle JohnsonSep 6, 2025
      If a command has been transmitted but nothing has happened, consider the following: * Has an error occurred? * Do you receive an answer message every time you send a setting command?
  • M
    mpriceSep 8, 2025
    What to do if data is garbled when transferring a file using a ":CARD:PICKout?" or ":MEMory:PICKout?" query on Hioki Measuring Instruments?
    • J
      James CampbellSep 8, 2025
      When data is garbled during file transfer, verify whether flow control is properly configured between the Hioki 3169-20/21 and the PC.
  • S
    sgloverSep 12, 2025
    What to do if the ":CARD" command does not function properly on Hioki 3169-20?
    • C
      Courtney RobinsonSep 12, 2025
      When ":CARD" command does not function properly, check the following: * Is the PC card properly installed? * Does the specified file exist?

Summary

Safety, Usage, and Setup

Pre-Use, Shipping, and Safety Precautions

Covers pre-use checks, shipping guidelines, safety notes, symbols, and measurement categories.

Usage Notes and Handling Guidelines

Provides precautions for operating the device, environmental considerations, and handling advice.

Instrument Overview and Parts

Product Overview and Key Features

General description and detailed list of main features including RS-232C, D/A output, and optional sensors.

Parts Identification and Screen Elements

Details instrument components (panels) and screen layout, display elements, and indicators.

Measurement Setup and Connections

Connection Procedures, Methods, and Ranges

Covers overall connection steps, wiring diagrams, range selection, and connection checks.

Configuration and Settings

Setting Screen Navigation and Measurement Settings

Covers accessing settings screens and configuring wiring, sampling, and ranges.

Data Output, Saving, and System Settings

Details settings for time-series measurement, data output, saving, and system parameters.