Why are setting commands rejected on my Hioki 3169-20 Measuring Instruments?
- DdleeAug 16, 2025
The settings cannot be changed while time-series measurement is on standby or in progress, or the measurement data is being held.
Why are setting commands rejected on my Hioki 3169-20 Measuring Instruments?
The settings cannot be changed while time-series measurement is on standby or in progress, or the measurement data is being held.
Why does the response message to a query differ from the display of the Hioki 3169-20/21 Measuring Instruments?
A response message is created after the Hioki 3169-20/21 has received a query. The message, therefore, may differ from the display when the PC reads the response message.
What to do if the number of read data is too small with Hioki 3169-20 Measuring Instruments?
If the data contains a comma, try using a LINE INPUT statement.
How to fix communication failure with RS-232C on Hioki 3169-20?
To resolve the communication failure, ensure that: * The message terminator (delimiter) of the PC is set correctly. * The PC settings match the RS-232C connection conditions set on the 3169-20/21.
What to do if the RS-232C bus hangs when reading data using BASIC INPUT statement on Hioki 3169-20 Measuring Instruments?
When attempting to read data using a BASIC INPUT statement and the RS-232C bus hangs, ensure that: * A query is transmitted before every INPUT#(ENTER). * None of the transmitted queries have resulted in an error.
How to troubleshoot Hioki 3169-20 Measuring Instruments when the RS-232C has stopped working completely?
To troubleshoot a completely non-functional RS-232C connection on your Hioki Measuring Instruments, check the following: * Are the cables properly connected? * Are all the devices powered on? * Are the correct cables being used? * Is the device to be connected to the RS-232C set to a PRINTER? If so, change the setting to a PC.
What to do if sending several queries produces only one response on Hioki Measuring Instruments?
When several queries produces only one response, consider the following: * Has an error occurred? * Send the queries one at a time, and read the responses individually. When you want to read them in all at once, try doing so by putting them all on one line separated by the message separator character.
What to do if nothing happens after transmitting a command to Hioki 3169-20 Measuring Instruments?
If a command has been transmitted but nothing has happened, consider the following: * Has an error occurred? * Do you receive an answer message every time you send a setting command?
What to do if data is garbled when transferring a file using a ":CARD:PICKout?" or ":MEMory:PICKout?" query on Hioki Measuring Instruments?
When data is garbled during file transfer, verify whether flow control is properly configured between the Hioki 3169-20/21 and the PC.
What to do if the ":CARD" command does not function properly on Hioki 3169-20?
When ":CARD" command does not function properly, check the following: * Is the PC card properly installed? * Does the specified file exist?