22
Power Measurement
Quick Start Manual
1. Press the key to display the measurement screen.
2. Press the (SCREEN) key to display the selection win-
dow.
3. Press the key to move the cursor to "WIRING CHECK".
4. Press the (select) key to display the wiring check
result.
The connection status is shown by the voltage, current vectors, and the con-
nection check result.
If the wiring check result is "NG," check the wiring.
The displayed voltage and current levels and phase angle are for the funda-
mental component. These values differ from RMS values.
Checking the Wiring
The wiring check result is NG. (p.23)