74
5.3 Setting on the Data Output Setting Screen (DATA OUTPUT)
5.3.3 Setting Interval
The data storable time varies depending on the setting of the
data storage items and interval.
The setting ranges for data output items vary depending on the
setting of the interval.
When the interval is 30 seconds or less, harmonic measure-
ment-data output and printer output are not available.
When a short-term interval is selected, the 3169-20/21 outputs
the instantaneous values of normal measurement only. The file
will be in binary format and must be converted to a text file to
be read into a generally available spreadsheet software. For
details on spreadsheet software, see the CD-R supplied with
the 3169-20/21.
Press the
SET UP
key to display the setting screen.
Press the
F5
(NEXT SCR) key to display the data output setting
screen. (3169-20: SET UP 2/4, 3169-21: SET UP 2/5)
Move the cursor to "
INTERVAL TIME
."
Press the
F1
(change) key to display the selection window.
Select an interval with the cursor key.
Press the
F1
(select) key.
Standard interval 1, 2, 5, 10, 15, 30 s, 1, 2, 5, 10, 15, 30, 60 m
Short-term interval Full wave (Each one cycle), 100m, 200m, 500ms
NEXT
SCR
change
select