5.2 Setting on the Measurement Setting Screen (MEASUREMENT)
67
1
2
3
4
5
6
7
8
9
10
11
5.2.9 Setting the Clamp-On Sensor
• When the wiring method is 3P4W4I, a clamp-on sensor can be
set for I4 that differs from that for I1 to I3.
• The range setting for the 9667 sensor is made on the sensor
(500/5000 A range). When power to the 9667 is turned off,
then back on again, the range is always set to 5000 A. When
the 9667 is to be used for an extended period with the AC
adapter, we recommend that a battery be used as well.
Press the
SET UP
key to display the setting screen.
Press the
F5
(NEXT SCR) key to display the measurement setting
screen. (3169-20: SET UP 1/4, 3169-21: SET UP 1/5)
Move the cursor to "
SENSOR
" of the circuit to be changed.
Press the
F1
(change) key to display the selection window.
Select a clamp-on sensor for each circuit by using the cursor key.
Press the
F1
(select) key.
9660 Use the 9660 CLAMP ON SENSOR (100 A rated).
9661 Use the 9661 CLAMP ON SENSOR (500 A rated).
9667-5kA Use the 9667 FLEXIBLE CLAMP ON SENSOR (5000 A rat-
ed). (5000 A range)
9667-500A Use the 9667 FLEXIBLE CLAMP ON SENSOR (500 A rat-
ed). (500 A range)
9669 Use the 9669 CLAMP ON SENSOR (1000 A rated).
9694 Use the 9694 CLAMP ON SENSOR (5 A rated).
9695-02 Use the 9695-02 CLAMP ON SENSOR (50 A rated).
9695-03 Use the 9695-03 CLAMP ON SENSOR (100 A rated).
NEXT
SCR
change