100
6.4 Measuring the Average, Maximum, and Minimum Values
Measures the average, maximum, and minimum values of the volt-
age, current, power, and harmonic
<Ending of Time-Series Measurement>
6.4 Measuring the Average, Maximum, and
Minimum Values
Set the parameters on the measurement setting, data output set-
ting, and save/print items setting screens.
5 Setting Procedure (page 55)
Press the
MEASURE
key to display the measurement screen.
Press the
START/STOP
key to start measurement. When the time-
series measurement has started, the LED lights up, indicating that
the 3169-20/21 is performing measurement. When the measure-
ment start method is set to "Time" or "Just", the 3169-20/21 will
stand by until the start time (the LED blinks) and start measure-
ment at the start time.
NEXT
SCR
If time-series measurement has been started by pressing the
START/STOP
key, the average, maximum, and minimum values
will be displayed on the screen.
The display shows the average, maximum, and minimum val-
ues of the measurements taken up to the current time from the
start of time-series measurement.
The pieces of data to be stored or printed out are the average,
maximum, and minimum values of every interval.
The average, maximum, and minimum values are not dis-
played for harmonic measurements.
The time-series measurement stops by the method set on the
data-output setting screen. Press the
START/STOP
key to stop the
measurement when the stoping method is set to manual, or stop it
immediately in any other mode.