6.4 Measuring the Average, Maximum, and Minimum Values
Measures the average, maximum, and minimum values of the volt-
age, current, power, and harmonic
<Ending of Time-Series Measurement>
6.4 Measuring the Average, Maximum, and
Minimum Values
Set the parameters on the measurement setting, data output set-
ting, and save/print items setting screens.
❖
5 Setting Procedure (page 55)
Press the
MEASURE
key to display the measurement screen.
Press the
START/STOP
key to start measurement. When the time-
series measurement has started, the LED lights up, indicating that
the 3169-20/21 is performing measurement. When the measure-
ment start method is set to "Time" or "Just", the 3169-20/21 will
stand by until the start time (the LED blinks) and start measure-
ment at the start time.
NEXT
SCR
• If time-series measurement has been started by pressing the
START/STOP
key, the average, maximum, and minimum values
will be displayed on the screen.
• The display shows the average, maximum, and minimum val-
ues of the measurements taken up to the current time from the
start of time-series measurement.
• The pieces of data to be stored or printed out are the average,
maximum, and minimum values of every interval.
• The average, maximum, and minimum values are not dis-
played for harmonic measurements.
The time-series measurement stops by the method set on the
data-output setting screen. Press the
START/STOP
key to stop the
measurement when the stoping method is set to manual, or stop it
immediately in any other mode.