3.4 Using a Clamp-On Sensor (Option)
29
1
2
3
4
5
6
7
8
9
10
Use Hioki clamp-on sensor 9660, 9661, 9667, 9669, 9694, or
9695-02/03.
3.4.1 Clamp-On Sensor Specifications
Refer to the Instruction Manual for the specific model for more
details.
9660 CLAMP ON SENSOR/ 9695-02 CLAMP ON SENSOR
3.4 Using a Clamp-On Sensor (Option)
Connect the clamp-on sensors to the product first, and then
to the active lines to be measured. Observe the following to
avoid electric shock and short circuits.
• Clamp sensor should only be connected to the secondary
side of a breaker, so the breaker can prevent an accident if
a short circuit occurs. Connections should never be made
to the primary side of a breaker, because unrestricted cur-
rent flow could cause a serious accident if a short circuit
occurs.
• When the clamp sensor is opened, do not allow the metal
part of the clamp to touch any exposed metal, or to short
between two lines, and do not use over bare conductors.
• To prevent damage to the product and sensor, never con-
nect or disconnect a sensor while the power is on.
• The current input terminals of the 3169-20/21 are not insu-
lated. To avoid the risk of electric shock, only use the spec-
ified optional clamp-on sensor.
9695-02
9660
Rated primary current 100 A AC
Output voltage 1 mVAC/A
Maximum permissible input 130 A continuous (at 45 to 66 Hz, 50
°
C)
Amplitude accuracy ±0.3%rdg. ±0.02%f.s. (f.s.=100 A, 45 to 66 Hz)
Phase accuracy Within ±1
°
(45 Hz to 5 kHz)
Amplitude frequency
characteristics
Within ±1% at 40 Hz to 5 kHz
(deviation from accuracy)
Maximum rated voltage to
earth
300 Vrms
Measurable conductor
diameter
15 mm max.
Operating temperature and
humidity
0 to 50
°
C (32
°
F to 122
°
F), 80% RH or less
(no condensation)
Option 9219 CONNECTION CABLE (for 9695-02/03)