3.4 Using a Clamp-On Sensor (Option)
31
1
2
3
4
5
6
7
8
9
10
9694 CLAMP ON SENSOR/ 9695-03 CLAMP ON SENSOR
3.4.2 Connecting a Clamp-on Sensor
9695-03
9694
Rated primary current 9694: 5 A AC/ 9695-03: 50 A AC
Output voltage 10 mVAC/A
Maximum permissible input 9694: 50 A continuous (at 45 to 66 Hz , 50
°
C)
9695-02: 60 A continuous (at 45 to 66 Hz , 50
°
C)
Amplitude accuracy 9694: ±0.3%rdg. ±0.02%f.s.(f.s.=5 A, 45 to 66 Hz)
9695-02: ±0.3%rdg. ±0.02%f.s.
(f.s.=5 A, 45 to 66 Hz)
Phase accuracy Within ±2
°
(45 Hz to 5 kHz)
Amplitude frequency
characteristics
Within ±1% at 40 Hz to 5 kHz
(deviation from accuracy)
Maximum rated voltage to
earth
300 Vrms
Measurable conductor
diameter
15 mm max.
Operating temperature and
humidity
0 to 50
°
C (32
°
F to 122
°
F), 80% RH or less
(no condensation)
Option 9219 CONNECTION CABLE (for 9695-02/03)
When disconnecting the BNC connector, be sure to release the
lock before pulling the connectors apart. Forcibly pulling the con-
nector without releasing the lock, or pulling on the cable, can
damage the connector.
Lock
BNC plug slots
1.
At each end of the cable, apply the
input cord labels having the same
color as the current input terminal to
which it is to be connected.
2.
Align the slots in the BNC plug with
the guide pins on the connector at
the instrument side, then push and
turn the plug clockwise.
(to unplug the connector, push the
plug and turn it counterclockwise
before pulling it apart.)
Connecting a Clamp Sensor to the Instrument
3169-20/21 Current
input terminal
connector guide
pins