64
5.2 Setting on the Measurement Setting Screen (MEASUREMENT)
5.2.6 Setting the VT Ratio (PT Ratio)
What is VT?
VT stands for "voltage transformer." It is also referred to as "PT"
(potential transformer). VT is used in high-voltage measurement to
convert (step-down) the voltage measured to a smaller level and
supply the conversion result to an instrument.
VT ratio (voltage transformation ratio): A ratio used to convert the
secondary voltage of VT to the primary voltage.
This ratio is used for measurement conducted on the secondary
side of an external voltage transformer (VT) to convert the volt-
age measured to the primary voltage to be displayed.
Press the
SET UP
key to display the setting screen.
Press the
F5
(NEXT SCR) key to display the measurement setting
screen. (3169-20: SET UP 1/4, 3169-21: SET UP 1/5)
Move the cursor to the "
VT
" digit to be changed.
Set the VT ratio by using the function keys as follows:
(Cursor : Moves left to next digit; Cursor : Moves right to next
digit)
Setting range: 0.01 to 9999.99 (Default: 1.00)
-
Decrements the number.
+
Increments the number.
-
+
NEXT
SCR