5.3 Setting on the Data Output Setting Screen (DATA OUTPUT)
69
1
2
3
4
5
6
7
8
9
10
11
5.3.1 Setting the Time-Series Measurement Start
Method
Press the
SET UP
key to display the setting screen.
Press the
F5
(NEXT SCR) key to display the data output setting
screen. (3169-20: SET UP 2/4, 3169-21: SET UP 2/5)
Move the cursor to "
MEAS. START."
Set the time-series measurement start method using the function
keys.
*
1
:
The 3169-20/21 will enter into waiting status when the START/STOP key is
pressed to start measurement. Measurement will then start at the set time.
MANUAL Manual Measurement starts when the START/STOP key is
pressed (default setting).
TIME Measurement starts at the set time.*1
JUST Measurement will begin as soon as the internal clock reaches
a time that is evenly divisible by the set interval.
Example:
If the current time is 10:41:22, and the measurement interval
is set to 30 minutes, then measurement will start at 11:00:00
(:00 minutes [or :60 minutes] is evenly divisible by 30 as
60/30 = 2 with no remainder) with the next recording at
11:30:00, 12:00:00, etc.). If the interval is set to 10 minutes,
measurement will start at 10:50:00.
NEXT
SCR
MANUAL
TIME
JUST
The
START/STOP
key is effective only when the measurement
screen is displayed. It is not effective when the setting screen is
displayed.