70
5.3 Setting on the Data Output Setting Screen (DATA OUTPUT)
Notes for long term measurement
Set the Time-Series Measurement Start Time
(when the start method is set to time).
The longest measurement period of this unit is one year.
The instrument will automatically stop time-series measure-
ment when one year has elapsed since the start of the time-
series measurement. To continue measurement, restart the
unit and then resume the time-series measurement.
The measurement settings are saved even after restarting
the unit.
Set the time again when using the "TIME" start and stop.
If the set measurement start time has already expired when the
START/STOP
key is pressed, the 3169-20/21 displays an error
message and starts measurement by the "Just" start method,
which commences measurement at the optimal time.
Move the cursor to the digit to be changed in the measurement
start date and time.
Set the start time using the function keys.
(Cursor : Moves left to next digit; Cursor : Moves right to next
digit)
-
Decrements the number.
+
Increments the number.
AUTO Set the start time to any subsequent time.
-
+
AUTO