The specifications below apply to the 3169-20/21 CLAMP ON POWER HiT-
ESTER.
Specifications
13
Environmental and Safety Specifications
Operating environment
Indoors, < 2000 m ASL (6562-ft.)
Storage temperature
and humidity
-10°C to 50°C (-14°F to 122°F), 80% RH or less (no condensation)
Operating temperature
and humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensating)
Dielectric strength
(50/60 Hz for 15 sec.)
5.55 kVrms (current sensitivity 1 mA)
Between voltage input terminals and instrument case
3.32 kVrms (current sensitivity 1 mA)
Between voltage input terminals and current input terminals,
between external interface terminals
2.3 kVrms (current sensitivity 1 mA)
Between power supply and instrument case
1.39 kVrms (current sensitivity 10 mA)
Between power supply and current input terminals,
between external interface terminals
Power supply
Rated power supply voltage: AC100 to 240 V
Rated power supply frequency: 50/60 Hz
(Voltage fluctuations of
±10% from the rated supply voltage are taken
into account.)
Maximum rated power
30 VA
Dimensions
Approx.
210W X 160H X 60D mm
(not including protrusions)
(8.27"W X 6.30"H X 2.36"D)
Mass
Approx.
1.2 kg (3169-20/21) (42.3 oz.)
Standards applying
Safety EN61010-1:2001
Pollution Degree 2, Measurement Category III
(anticipated transient overvoltage 6000V)
EMC EN61326:1997+A1:1998+A2:2001+A3:2003 Class A
EN61000-3-2:2000
EN61000-3-3:1995+A1:2001