EasyManua.ls Logo

Hioki 3169-20 - Page 178

Hioki 3169-20
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
170
Operations after Reset
3169-20/21 Outage during standby state
If the instrument is reset before the set time-
series measurement start time, the instru-
ment returns to the standby state. It starts
time-series measurement at the set start
time.
If the instrument is reset after the set time-
series measurement start time, the instru-
ment starts time-series measurement at the
next interval time.
Outage during time-series measurement
The time-series measurement is suspended
during the outage. After the instrument is
reset, it performs time-series measurement
for the remaining time. The measurements
during the outage will be treated as "0."
PC Card/
Internal Memory
If a power outage occurs while the memory is
being accessed, the data being saved may
be lost or, in the worst case, the file may be
broken. If the outage occurs during time-
series measurement, the outage time and
reset time will be saved after the reset.
If the memory is not accessed during the out-
age and the outage occurs during time-series
measurement, the outage time and reset time
will be saved after the reset.
Printer If a power outage occurs during printing, the
printer stops printing immediately (except when
the printer is operating on battery power).

Table of Contents

Other manuals for Hioki 3169-20

Related product manuals