154
10.2 Functions of the External Input/Output Terminal
Quick Start Manual
Output terminal
Status Output
Outputs a status signal indicating that the 3169-20/21 is performing
time-series measurement.
During time-series measurement: Short (low level)
Status other than time-series measurement: Open (high level)
Output signal: Open-collector output (with voltage out-
put)
Output-voltage range: High level: 4.5 V to 5.0 V
Low level: 0 V to 0.5 V
Maximum input voltage: 0 to 30 V, 50 mAmax., 200 mWmax.
+5 V
Output terminal
GND
10 k
When the 3169-20/21 is standing by for measurement, it is
treated as status other than time-series measurement.
The lag of a signal is approximately 600 ms.