Specications
Specications
Accuracy
We dene measurement tolerances in terms of rdg. (reading)
and dgt. (digit) values, with the following meanings:
rdg.
(reading or
displayed value)
The value currently being measured and
indicated on the measuring instrument.
dgt.
(resolution)
The smallest displayable unit on a digital
measuring instrument, i.e., the input
value that causes the digital display to
show a “1” as the least-signicant digit.
General Specications
Operating
environment
Indoors, pollution degree 2, altitude up
to 2000 m (6562 ft.)
Operating temperature and humidity
Temperature −25°C to 65°C (−13.0°F to 149.0°F)
(From the 40 M
Ω
range: up to 40°C)
Humidity
(no
condensation)
Less than 40°C (104.0°F): 80% RH or less
At least 40°C (104.0°F) but less than 45°C
(113.0°F): 60% RH or less
At least 45°C (113.0°F) but less than 50°C
(122.0°F): 50% RH or less
At least 50°C (122.0°F) but less than 55°C
(131.0°F): 40% RH or less
At least 55°C (131.0°F) but less than 60°C
(140.0°F): 30% RH or less
At least 60°C (140.0°F) but less than 65°C
(149.0°F): 25% RH or less