EasyManua.ls Logo

Hioki DT4254 - Page 93

Hioki DT4254
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
85
Accuracy Table
Minimum sensitivity voltage (sine wave)
Range Measurement range
AC voltage range
6.000 V 60.00 V 600.0 V 1000 V
99.99
Hz
5.00 Hz to 99.99 Hz
*1
0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
999.9
Hz
100.0 Hz to 999.9 Hz 0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
9.999
kHz
1.000 kHz to 9.999 kHz 0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
99.99
kHz
10.00 kHz to 50.00 kHz 1.800 V
or more
12.00 V
or more
120.0 V
or more
230 V or
more
Over 50.00 kHz to 99.99
kHz
3.000 V
or more
24.00 V
or more
240.0 V
or more
400 V or
more
The voltage input is up to 2×10
7
V • Hz.
[----] appears when no measurement can be made.
*1: The measurement range from 5.00 Hz is only for the 6.000 V range. The
measurement range for other voltage ranges is 40.00 Hz to 99.99 Hz.
Minimum sensitivity current (sine wave)
Range Measurement range
AC current range
600.0 mA 6.000 A 10.00 A
99.99 Hz 40.00 Hz to 99.99 Hz 60.0 mA
or more
0.600 A or
more
3.00 A or
more
999.9 Hz 100.0 Hz to 999.9 Hz 60.0 mA
or more
0.600 A or
more
3.00 A or
more
9.999 kHz 1.000 kHz to 9.999 kHz 60.0 mA
or more
0.600 A or
more
3.00 A or
more