DIGITAL MULTIMETER DT4200 Series
New wireless “Middle Model” DT4261 between the Standard
and High-end models of the DT4200 lineup!

Questions and Answers

  • D
    dennispatrickAug 9, 2025
    What to do if nothing appears on my Hioki Multimeter display?
    • L
      larryparksAug 9, 2025
      First, check that the batteries are not exhausted and replace them with new ones. Also, verify that the auto power save function has not been activated and check its settings.
  • K
    Karen LopezAug 14, 2025
    What to do if my Hioki DT4200 is not communicating with PC?
    • V
      vanessathompsonAug 15, 2025
      Check the communication settings between the instrument and the PC to ensure they are correct. Verify that the baud rate and parity check are set correctly. Ensure the USB cable is correctly connected. Also, confirm that the light receiving and emitting parts are clean.
  • S
    Stephanie GrossAug 25, 2025
    What to do if Hioki DT4200 display shows '-----'?
    • N
      Nicolas TurnerAug 25, 2025
      The '----' appears when the rotary switch position is not confirmed. Set the rotary switch to the proper position.
  • J
    Jason CarrilloAug 29, 2025
    Why does the Hioki DT4200 display fluctuate?
    • M
      mitchellemilyAug 29, 2025
      Ensure that the input signal is within the input range for the instrument. If noise is affecting the reading, use the filter function of the instrument.
  • J
    Jennifer LeeSep 3, 2025
    What to do if turning on Hioki DT4200 Multimeter brings up error display?
    • M
      Matthew WalkerSep 3, 2025
      Reset the instrument. If the same symptom still occurs even after resetting the instrument, send the instrument for repair.
  • S
    Scott OneillSep 10, 2025
    Why Hioki DT4200 Multimeter measurement value does not appear?
    • P
      Philip BlackSep 12, 2025
      If the measured current value does not appear, check if the fuse is blown and replace it with the specified fuse if needed. Ensure that the fuse holder is not deformed; if it is, restore its shape using needle-nose pliers. Check that the test lead is not broken by performing a continuity check and replace the lead if necessary. Also, verify that the test leads have been properly inserted. Finally, confirm that the measurement method being used is correct.

Hioki DT4200 Series Specifications

General IconGeneral
TypeDigital Multimeter
Temperature MeasurementThermocouple (K type)
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Capacitance, Frequency, Temperature
DC Voltage Range600mV to 1000V
DC Current Range600μA to 10A
AC Current Range600μA to 10A
Resistance Range600Ω to 60MΩ
InterfaceUSB
Power SupplyAA batteries (x4)
WeightApprox. 340g (including batteries)

Related product manuals