General Specications
Durability
Drop proof Yes
Operating temperature and humidity*
1
-15°C to 55°C
Storage temperature and humidity*
2
-30°C to 60°C
Applicable standards
Safety: EN61010, EMC: EN61326; Waterproof and dustproof: IP40
Accessories
Safety
Maximum rated voltage
between
input terminals and ground
CAT Ⅲ 1000 V, CAT Ⅳ 600 V
Maximum rated voltage
between terminals
Between the V and COM terminals: 1000 V DC/AC
Maximum rated current
between terminals
Between the mA and COM terminals: 600 mA DC/600 mA AC
Between the A and COM terminals: 10 A DC/10 A AC
TEST LEAD L9207-10, Instruction Manual, LR6 alkaline battery × 4
93W × 197H × 53D mm (3.66″W × 7.76″H × 2.09″D),
650 g (23 oz.) (including batteries)
Dimensions / Weight
*1. -15°C to 55°C (5°F to 131°F), Up to 40°C (104°F): at 80% RH or less (non-condensating),
40°C to 45°C (104°F to 113°F): at 60% RH or less (non-condensating),
45°C to 55°C (113°F to 131°F): at 50% RH or less (non-condensating)
*2. 80%RH or less (non-condensating)
High-End
Middle
DT4261
(Accuracy guaranteed for 1 year, post-adjustment accuracy guaranteed for 1 year)
DC Voltage
Range Accuracy
※1
Input Impedance
600.0 mV ±0.15% rdg. ±5 dgt.
11.3 MΩ ± 2.0 %
6.000 V
±0.15% rdg. ±2 dgt.60.00 V 10.4 MΩ ± 2.0 %
600.0 V
10.3 MΩ ± 1.5 %
1000 V ±0.15% rdg. ±5 dgt.
Resistance Measurement
Range Accuracy
Measurement Current
Open-terminal Voltage
600.0 Ω ±0.7% rdg. ±5 dgt. Approx. 200 μA
DC 2.0 V or less
6.000 kΩ
±0.7% rdg. ±3 dgt.
Approx. 100 μA
60.00 kΩ
Approx. 10 μA
600.0 kΩ
Approx. 1 μA
6.000 MΩ
±0.9% rdg. ±3 dgt.
Approx. 100 nA
60.00 MΩ
±1.5% rdg. ±3 dgt.
Approx. 10 nA
Accuracy guarantee condition After zero adjustment has been performed
Continuity Check
Range Accuracy
Measurement Current
Open-terminal Voltage
600.0 Ω
±0.7% rdg. ±5 dgt.
Approx. 200 μA DC 2.0 V or less
Continuity ON threshold
Approx. 25 Ω or less (continuous buzzer sound, red backlight on)
Continuity OFF threshold
Approx. 245 Ω or more (buzzer sound off, red backlight off)
Diode Check
Range Accuracy
Measurement Current
Open-terminal Voltage
1.800 V
±0.5% rdg. ±5 dgt.
Approx. 200 μA DC 2.0 V or less
Forward threshold
Intermittent buzzer sound at 0.15 V to 1.8 V, continuous buzzer
sound at less than 0.15 V, red backlight on.
AC Clamp (AC Current)
Range
Accuracy
40 Hz to 500 Hz 500 Hz to 1 kHz
10.00 A
±0.9% rdg. ±3 dgt. ±1.5% rdg. ±3 dgt.
20.00 A
50.0 A
100.0 A
200.0 A
500 A
1000 A
The optional 9010-50, 9018-50, or 9132-50 CLAMP ON PROBE is used.
Accuracy does not include the error of the clamp-on probe.
Crest factor 3 or less
Accuracy specication range
Minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range
Capacitance Measurement
Range Accuracy Measurement Current
Open-terminal Voltage
1.000 μF
±1.9% rdg. ±5 dgt.
Approx. 10 n,100 n,1 μA
DC 2.0 V or less
10.00 μF Approx. 100 n,1 μ,10 μA
100.0 μF
Approx. 1 μ,10 μ,100 μA
1.000 mF
Approx. 10 μ,100 μ,200 μA
10.00 mF ±5.0% rdg. ±20 dgt. Approx. 100 μ,200 μA
Frequency
Range Accuracy
99.99 Hz
±0.1% rdg. +1 dgt.
999.9 Hz
9.999 kHz
99.99 kHz (V AC Only)
AC Voltage
Range
Accuracy
Input Impedance
40 Hz to 500 Hz 500 Hz to 1 kHz
6.000 V
±0.9% rdg. ±3 dgt. ±1.5% rdg. ±3 dgt.
11.3 MΩ ± 2.0% // 100 pF or less
60.00 V
10.4 MΩ ± 2.0% // 100 pF or less
600.0 V
10.3 MΩ ± 1.5% // 100 pF or less
1000 V
DC A Measurement
Range Accuracy Input Impedance
600.0 mA
±0.5% rdg. ±3 dgt. 35 mΩ ±30%6.000 A
10.00 A
AC A Measurement
Range
Accuracy
Input Impedance
40 Hz to 500 Hz 500 H to 1 kHz
600.0 mA
±1.4% rdg. ±3 dgt. ±1.8% rdg. ±3 dgt. 35 mΩ ±30%6.000 A
10.00 A
Crest factor
3 at up to 4000 counts and reduces linearly to 2 at 6000 counts.
1000 V range only: 2 at up to 750 counts, linearly decreasing to
1.5 at 1000 counts.
Accuracy specication range
For ACV, minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range.
Crest factor 3 at up to 4000 counts and reduces linearly to 2 at 6000 counts.
Accuracy specication range
For ACV, minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range.
Accuracy specication range
Add ±2 dgt. when measuring at or below 5% of range.
*1. Add ±1 dgt. when measuring at or below 5% of range