Hioki DT4200 Series - Page 15

Hioki DT4200 Series
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
High-End
DT4281 / DT4282
AC Voltage
Range
Accuracy
20 Hz to 45 Hz 45 Hz to 65 Hz 65 HZ to 1 kHz 1 kHz to 10 kHz 10 kHz to 20 kHz 20 kHz to 100 kHz
60.000 mV
±1.3% rdg.
±60 dgt.
±0.4% rdg.
±40 dgt.
±0.% rdg.
±40 dgt.
±0.9% rdg.
±40 dgt.
±1.5% rdg.
±40 dgt.
±20% rdg. ±80 dgt.
600.00 mV
±8% rdg. ±80 dgt.
6.0000 V
±1% rdg.
±60 dgt.
±0.2% rdg.
±25 dgt.
±0.3% rdg.
±25 dgt.
±0.4% rdg.
±25 dgt.
±0.7% rdg.
±40 dgt.
±3.5% rdg.
±40 dgt.
60.000 V
Undened
600.00 V
Undened Undened
1000.0 V
DC V + AC V Measurement
Range
Accuracy
20 Hz to 45 Hz 45 Hz to 65 Hz 65 HZ to 1 kHz 1 kHz to 10 kHz 10 kHz to 20 kHz 20 kHz to 100 kHz
6.0000 V
±1.2% rdg.
±65 dgt.
±0.3% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±1.5% rdg.
±45 dgt.
±3.5% rdg.
±125 dgt.
60.000 V
Undened
600.00 V
Undened Undened
1000.0 V
±0.4% rdg.
±45 dgt.
Input impedance
1 MΩ ±4% // 100 pF or less
Crest factor 3 or less (1.5 or less for the 1000.0 V range)
Accuracy
specication range
5% or more of each range
With the lter ON, accuracy is dened only for frequencies
100 Hz or less. Furthermore, 2% rdg. is added.
DC A Measurement
*1. DT4282 only
Range
Accuracy / Display update : slow
Accuracy / Display update : normal
Shunt Resistance
600.00 μA
±0.05% rdg. ±5 dgt.
±0.05% rdg. ±25 dgt.
101 Ω
6000.0 μA ±0.05% rdg. ±5 dgt.
60.000 mA ±0.05% rdg. ±25 dgt.
1 Ω
600.00 mA ±0.15% rdg. ±5 dgt. ±0.15% rdg. ±5 dgt.
6.0000 A
*1
±0.2% rdg. ±5 dgt.
±0.2% rdg. ±25 dgt.
10m Ω
10.000 A
*1
±0.2% rdg. ±5 dgt.
AC A Measurement
*1. DT4282 only
Range
Accuracy
20 Hz to 45 Hz 45 Hz to 65 Hz 65 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 20 kHz
600.00 μA
±1.0% rdg.
±20 dgt.
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±20 dgt.
±2% rdg.
±20 dgt.
±4% rdg.
±20 dgt.
6000.0 μA
±1.0% rdg.
±5 dgt.
±0.6% rdg.
±5 dgt.
±0.6% rdg.
±5 dgt.
±2% rdg.
±5 dgt.
±4% rdg.
±5 dgt.
60.000 mA
±1.0% rdg.
±20 dgt.
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±20 dgt.
±1% rdg.
±20 dgt.
±2% rdg.
±20 dgt.
600.00 mA
±1.0% rdg.
±5 dgt.
±0.6% rdg.
±5 dgt.
±0.6% rdg.
±5 dgt.
±1.5% rdg.
±10 dgt.
Undened
6.0000 A
*1
Undened
±0.8% rdg.
±20 dgt.
±0.8% rdg.
±20 dgt.
Undened Undened
10.000 A
*1
Undened
±0.8% rdg.
±5 dgt.
±0.8% rdg.
±5 dgt.
Undened Undened
Shunt resistance μA Range 101 Ω, mA Range 1Ω, A Range 10 mΩ
Crest factor 3 or less (Note that it applies to 1/2 of the range.)
Accuracy specication range
Accuracy is not dened for measurements below 5% of range
Continuity Check
Range
Accuracy
Measurement Current
Open-terminal Voltage
600.0 Ω
±0.5% rdg. ±5 dgt.
640 μA ±10% DC 2.5 V or less
Continuity threshold
20 Ω (default), 50 Ω, 100 Ω, 500 Ω
Capacitance Measurement
Range Accuracy
Measurement Current Open-circuit Voltage
1.000 nF
±1% rdg. ±20 dgt.
32 μA ±10% DC 2.5 V or less
10.00 nF
±1% rdg. ±5 dgt.
100.0 nF
1.000 μF
10.00 μF
±2% rdg. ±5 dgt.
680 μA ±20%
DC 3.1 V or less
100.0 μF
1.000 mF
DC 2.1 V or less10.00 mF
100.0 mF
±2% rdg. ±20 dgt.
Resistance Measurement
Range Accuracy
Measurement Current
Open-terminal Voltage
60.000 Ω
±0.3% rdg. ±20 dgt.
640 μA ±10%
DC 2.5 V or less
600.00 Ω
±0.03% rdg. ±10 dgt.
6.0000 kΩ
±0.03% rdg. ±2 dgt.
96 μA ±10%
60.000 kΩ
9.3 μA ±10%
600.00 kΩ
0.96 μA ±10%
6.0000 MΩ
±0.15% rdg. ±4 dgt.
96 nA ±10%
60.00 MΩ
±1.5% rdg. ±10 dgt.
600.0 MΩ
±3.0% rdg. ±20 dgt.
±8.0% rdg. ±20 dgt.
AC Clamp (AC Current)
DT4281 only
Range
Accuracy
40 Hz to 65 Hz 65 Hz to 1 kHz
10.00 A ±0.6% rdg. ±2 dgt. ±0.9% rdg. ±2 dgt.
20.00 A ±0.6% rdg. ±4 dgt. ±0.9% rdg. ±4 dgt.
50.00 A ±0.6% rdg. ±10 dgt. ±0.9% rdg. ±10 dgt.
100.0 A ±0.6% rdg. ±2 dgt. ±0.9% rdg. ±2 dgt.
200.0 A ±0.6% rdg. ±4 dgt. ±0.9% rdg. ±4 dgt.
500.0 A ±0.6% rdg. ±10 dgt. ±0.9% rdg. ±10 dgt.
1000 A ±0.6% rdg. ±2 dgt. ±0.9% rdg. ±2 dgt.
DC Voltage
Range Accuracy Input Impedance
60.000 mV ±0.2% rdg. ±25 dgt.
1 GΩ or more // 100 pF or less
600.00 mV ±0.025% rdg. ±5 dgt.
6.0000 V
±0.025% rdg. ±2 dgt.
11.0 MΩ ±2% // 100 pF or less
60.000 V 10.3 MΩ ±2% // 100 pF or less
600.00 V
±0.03% rdg. ±2 dgt. 10.2 MΩ ±2% // 100 pF or less
1000.0 V
The optional 9010-50, 9018-50, or 9132-50 CLAMP ON PROBE is used.
Accuracy does not include the error of the clamp-on probe.
Crest factor 3 or less
Accuracy is not dened for measurements below 15% of range
Conductance (nS)
DT4282 only
Range Accuracy
Measurement Current Open-circuit Voltage
600.00 nS
±1.5% rdg. ±10 dgt.
96 nA ±10% DC 2.5 V or less
Accuracy is dened for humidity 60% RH or less. Accuracy is dened for the range
20nS or more. In the case of 300 nS or more, ±20 dgt. is added.
Diode Check
Range Accuracy
Measurement Current
Open-terminal Voltage
3.600 V
±0.1% rdg. ±5 dgt.
1.2 mA or less DC 4.5 V or less
Forward threshold
0.15 V, 0.5 V (default), 1 V, 1.5 V, 2 V, 2.5 V, 3 V
If the reading is lower than the threshold during the forward con-
nection, a buzzer sounds and the red backlight turns on.
Peak Measurement
(For AC V, DC V, DC+AC V, Clamp, DC μA, DC mA, DC A, AC μA, AC mA, AC A)
Main measurement
Signal width Accuracy
DC V
4 ms or more (single) ±2.0% rdg. ±40 dgt.
1 ms or more (repeated) ±2.0% rdg. ±100 dgt.
Other than
DC V
1 ms or more (single) ±2.0% rdg. ±40 dgt.
250 μs or more (repeated) ±2.0% rdg. ±100 dgt.
4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000, 1200 Ω
(default: 600 Ω)
Temperature
Thermocouple Type Range Accuracy
K -40.0 to 800.0 °C (-40.0 to 1472.0°F)
±0.5% rdg. ±3°C (5.4°F)
The optional K Thermocouple DT4910 is used. Accuracy does not include the error of the K thermocouple.
Frequency (For AC V, DC + AC V, AC μA, AC mA, AC A)
Range Accuracy
99.999 Hz
±0.005% rdg. +3 dgt.
999.99 Hz
9.9999 kHz
99.999 kHz
±0.005% rdg. +3 dgt.
500.00 kHz
Decibel Conversion Measurement : Standard impedance (dBm)
Measurement range
0.5 Hz or more ([----] is displayed when frequency is less than 0.5 Hz)
Pulse width 1 μs or more (DUTY ratio is 50%)
With the lter ON, accuracy is dened only for frequencies 100 Hz or less. (For ACV, DC+ACV)
(Accuracy guaranteed for 1 year, post-adjustment accuracy guaranteed for 1 year)