91
Accuracy Table
• Connection method: DC coupling
*1: The accuracy is specied in 1% or more of the range, however, ±5 dgt.
should be added to 5% or less of the range.
• Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured
values outside the accuracy guarantee range for frequency are also
displayed.) The accuracy is not specied for strain waveforms outside
the range of 40 Hz to 1 kHz.
• For 100 Hz with the lter ON, ±1.5% rdg. is added to the accuracy
specication between 40 Hz and 100 Hz and the accuracy is not specied
in 100 Hz or more.
• For 500 Hz with the lter ON, ±0.5% rdg. is added to the accuracy
specication between 40 Hz and 500 Hz and the accuracy is not specied
in 500 Hz or more.
12
DC Current (µA)
Range Accuracy Input impedance
60.00 µA ±0.8% rdg. ±5 dgt. 1 k
Ω
±5%
600.0 µA ±0.8% rdg. ±5 dgt. 1 k
Ω
±5%
• Overload protection: 250 mA/1000 V fuse, breaking capacity 50 kA AC/30 kA DC
• Auto range movement threshold: 6000 counts or more for upper range, 540
counts or less for lower range.
13
DC Current (mA)
Range Accuracy Input impedance
6.000 mA ±0.8% rdg. ±5 dgt. 15
Ω
±40%
60.00 mA ±0.8% rdg. ±5 dgt. 15
Ω
±40%
• Overload protection: 250 mA/1000 V fuse, breaking capacity 50 kA AC/30 kA DC
• Auto range movement threshold: 6000 counts or more for upper range, 540
counts or less for lower range.