91
Accuracy Table
Connection method: DC coupling
*1: The accuracy is specied in 1% or more of the range, however, ±5 dgt.
should be added to 5% or less of the range.
Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured
values outside the accuracy guarantee range for frequency are also
displayed.) The accuracy is not specied for strain waveforms outside
the range of 40 Hz to 1 kHz.
For 100 Hz with the lter ON, ±1.5% rdg. is added to the accuracy
specication between 40 Hz and 100 Hz and the accuracy is not specied
in 100 Hz or more.
For 500 Hz with the lter ON, ±0.5% rdg. is added to the accuracy
specication between 40 Hz and 500 Hz and the accuracy is not specied
in 500 Hz or more.
12
DC Current (µA)
Range Accuracy Input impedance
60.00 µA ±0.8% rdg. ±5 dgt. 1 k
Ω
±5%
600.0 µA ±0.8% rdg. ±5 dgt. 1 k
Ω
±5%
Overload protection: 250 mA/1000 V fuse, breaking capacity 50 kA AC/30 kA DC
Auto range movement threshold: 6000 counts or more for upper range, 540
counts or less for lower range.
13
DC Current (mA)
Range Accuracy Input impedance
6.000 mA ±0.8% rdg. ±5 dgt. 15
Ω
±40%
60.00 mA ±0.8% rdg. ±5 dgt. 15
Ω
±40%
Overload protection: 250 mA/1000 V fuse, breaking capacity 50 kA AC/30 kA DC
Auto range movement threshold: 6000 counts or more for upper range, 540
counts or less for lower range.