85
Accuracy Table
Current
measurement
range
Peak measurement range
DCA/ACA
600.00 µA 1200.0 µA
*2: Up to 600 mA for the
RMS
*3: Up to 10 A for the
RMS
6000.0 μA 12000 μA
60.000 mA 120.00 mA
600.00 mA 1200.0 mA
*2
6.0000 A (DT4282) 12.000 A
*3
10.000 A (DT4282) 15.00 A
*3
AC clamp
measurement
range (DT4281)
Peak measurement range
10.00 A 30.00 A
20.00 A 60.00 A
50.00 A 150.0 A
100.0 A 300.0 A
200.0 A 600.0 A
500.0 A 1500 A
1000 A 3000 A
5
Continuity check
Range Accuracy Measurement current
600.0
Ω
±0.5% rdg. ±5 dgt. 640 µA ±10%
Open circuit voltage 2.5 V DC or less
Continuity threshold 20
Ω
(default) /50
Ω
/100
Ω
/500
Ω
Threshold setting tolerance ±1% setting ±0.5
Ω
Response time Open circuit or short circuit is detected for at
least 10 ms.