  • K
    Kelsey MorganAug 7, 2025
    Why do the measurement values differ on my Hioki IM3523A when measuring a known test sample?
    • J
      Jeff ClarkAug 7, 2025
      The measurement values may differ because the measurement conditions of the known test sample and the instrument do not match. Ensure that the measurement conditions match. Also, perform open and short compensation again. Turn load compensation off. Ensure there is an appropriate trigger delay and trigger synchronization output wait time. If you are measuring a test sample other than a capacitor, turn the DC bias function OFF.
  • J
    Jessica GarciaAug 12, 2025
    What to do if keys do not work on Hioki IM3523A Measuring Instruments?
    • C
      christopher77Aug 12, 2025
      First, check if the keys are locked and disable the key lock. Second, check if the unit is being remotely operated from an external device using the communication cable and switch to the local state.
  • E
    Edwin RodriguezAug 16, 2025
    Why are key response and screen drawing slow on my Hioki Measuring Instruments?
    • J
      Jessica SantiagoAug 16, 2025
      When the measurement value automatic output function is enabled, key response and screen drawing may become slow in order to give priority to measurement and measurement value output.
  • A
    Abigail CarlsonAug 20, 2025
    Why is an error beep sound emitted continuously from my Hioki Measuring Instruments?
    • B
      Benjamin MilesAug 20, 2025
      When the measurement value automatic output function is enabled, a send error occurs on the measuring instrument side if the receive operation is not performed on the PC side, and a send error sound is emitted continuously when, for example, there is an internal trigger. Perform the receive operation on the PC side and then perform measurement on the measuring instrument side, or disable the measurement value automatic output function.
  • D
    Dean MartinezAug 25, 2025
    Why is a contact error generated on Hioki Measuring Instruments even though the connections are correct?
    • J
      Julia MillerAug 25, 2025
      If you are measuring an element other than a capacitor using the DC bias function, turn the DC bias function OFF.
  • G
    greengaryAug 28, 2025
    What to do if a key other than the pressed one is pressed on Hioki Measuring Instruments?
    • R
      Ryan FosterAug 28, 2025
      Perform panel compensation.
  • K
    Kristi MillerSep 2, 2025
    Why does the LCD appear blurred on Hioki IM3523A Measuring Instruments?
    • O
      obarajasSep 2, 2025
      If you are pressing the LCD screen too hard, press it gently. Slight blurring may occur, but this is normal.