The Test will end when the Target Voltage is reached and the Dwell Time reaches 0:00
or when the Max Leakage current limit is reached.
To turn off High Voltage at any time, press STOP, engage the E-STOP, or open the
Interlock Circuit.
MANUAL Test with Digital Meter Display
Setting up a Manual Test
1. Input Overvoltage (0.1→ 80kV)
2. Input Overcurrent (0.1 → 10,000µA or 0-10mA)
3. Confirm GST or UST mode
4. Press HV ON
5. Turn Rotary Knob or touch Up-Arrow button to activate and adjust High Voltage
To turn off High Voltage at any time, press STOP, engage the E-STOP, or open the
Interlock Circuit .
Note: When performing withstand tests on cable or capacitive loads, Manual mode is
recommended. At the
conclusion of the test, ramp the voltage back down to Zero before pressing
STOP.